You are viewing the site in preview mode
Skip to main content
|
Tonometer/position
|
Mean IOP, mean ± SD (range), mmHg
|
|---|
|
Sitting
|
|
GAT
|
14.5 ± 2.9 (5 to 22)
|
|
IcarePRO
|
13.3 ± 3.2 (6.3 to 23.6)*
|
|
IC200-single
|
11.6 ± 3.2 (3.6 to 21.2)*†
|
|
I200-continuous
|
11.5 ± 3.2 (6.5 to 25.3)*†
|
|
Supine
|
|
IcarePRO
|
14.5 ± 3.8 (6.5 to 25.3)
|
|
IC200-single
|
14.2 ± 4.4 (3.2 to 26.7)
|
|
IC200-continuous
|
14.3 ± 4.5 (3.4 to 29.5)
|
- Tukey’s multiple-comparison test
- GAT Goldmann applanation tonometry; IOP intraocular pressure; SD standard deviation
- *P < 0.05 as compared with GAT